Issues in 10.1016/j.inoche.2026.116862
Several issues are found in this article [1]:
EDX peaks in Figure 1 (c) are mislabeled: The peak around 0.376 keV is assigned to C Kα; however, the expected peak for C Kα lines around 0.277 keV. The peak around 0.451 keV is assigned to O Kα; however, the expected peak for C Kα lines around 0.525 keV. The peak around 4.643 keV is assigned to Ti Kα; however, the expected peak for C Kα lines around 4.512 keV. And the expected Ti Kβ is missing in this spectrum.
The manuscript states that "Additional reflections are observed at 17.3o and 22.6o"; however, no peak around 22.6o is observed in Figure 1 (b).
Figure 3 and Figure 5: MIC and MBC are discrete thresholds obtained by two-fold serial dilution and can only take values such as 64, 128 or 256 µg/mL. Reporting them as mean ± SD (e.g., 128 ± 6 µg/mL) is therefore not statistically meaningful: with n = 3, any non-zero SD would be at least ~37 µg/mL (e.g., 128, 128, 64), and a mean of exactly 128 with SD 6 cannot arise from the two-fold grid. Moreover, an arithmetic mean such as 96 µg/mL (from 64 and 128) is not a tested concentration and has no MIC interpretation.