Mislabeled EDX Peaks in 10.1016/j.inoche.2024.113092


Several peak assignments in Figure 5 are inconsistent with the characteristic X-ray energies. The peaks labeled C appear at approximately 0.17 keV in panel (a) and 0.05 keV in panel (b), whereas the C Kα line lies at 0.277 keV. The peaks labeled N lie at 1.26 - 1.34 keV in panel (a) and 1.43 - 1.48 keV in panel (b), far from the N Kα line at 0.392 keV. The peaks labeled Cu lie at 2.16 - 2.19 and 2.22 - 2.25 keV, whereas the Cu Lα line lies at 0.930 keV and the Cu Kα line at 8.041 keV falls outside the 0 - 5 keV window. In contrast, the O Kα line (0.525 keV) is correctly observed at 0.52 - 0.53 keV.



Reference

[1] 10.1016/j.inoche.2024.113092