Mislabeled EDX Peaks in 10.3389/fmats.2025.1602816


Several peak assignments in Figure 4 are inconsistent with the characteristic X-ray energies. The peaks labeled C appear at approximately 1.7 keV, whereas the C Kα line lies at 0.277 keV. The peaks labeled N lie at 1.3 keV, far from the N Kα line at 0.392 keV. The peaks labeled Ru lie at 2.2 keV, whereas the Ru line lies at 2.558 keV. In contrast, the O Kα line (0.525 keV) is correctly observed at 0.54 keV.